4. To add a new segment:
a. Tap the row containing the start time of the new segment.
b. Tap the End Time field and enter the start time of the new segment as the
end time of this segment.
c. Change the basal rate, if necessary.
d. Tap NEXT.
e. Then define the end time and basal rate for any following segments, as
needed.
5. To delete a segment:
a. Note the end time of the segment you want to delete.
b. Tap the segment that precedes the segment you want to delete.
c. Tap the End Time field and enter the end time of the segment you want to
delete. This "overwrites" the segment you want to delete.
d. Tap NEXT.
e. Then define the end time and basal rate for any following segments, as
needed.
6. When the Basal Program is correct, tap SAVE.
Note: If the basal rate for a segment is 0 U/hr, the PDM displays a message calling
this to your attention. Tap OK if the 0 U/hr rate is correct. Otherwise, tap CANCEL
and edit the segment with the 0 U/hr rate.
Note: To create additional Basal Programs after setup is complete, see page 75.

Temp basal configuration

For a description of temporary basal rates, also called temp basals, see page 138.
1. If you want the ability to use temp basals, tap the toggle to the "on" position.
The toggle is in the "on" position when it is in the right-hand position and is
blue.
If you turn temp basals on, percentages are used by default. To specify temp
basals as a flat rate (U/hr), see "Temp basal" on page 105.
2. Tap NEXT.