Pod Specifications - Insulet Omnipod Dash User Manual

Insulin management system
Table of Contents

Advertisement

Duration of Insulin
Action
Maximum bolus size
Extended Bolus
Bolus preset
Suspend
Low reservoir volume
advisory
Pod expiry notification 1 to 24 hours in 1-hour increments. Default is 4 hours.
Auto-off timer
History screen display
Language

Pod Specifications

Size: 3.9 cm wide x 5.2 cm long x 1.45 cm high (1.53" x 2.05" x 0.57")
Weight (without insulin): 26 grams (0.92 oz)
Operating temperature range: Pod operating environment of 5 °C to 40 °C
(41 °F to 104 °F).
Note: The Pod temperature equilibrates from 23 °C to 37 °C (73 °F to 98.6 °F)
when worn on the body.
Start-up temperature: above 10 °C (50 °F)
Storage temperature range: 0 °C to 30 °C (32 °F to 86 °F)
Reservoir volume (deliverable): 200 units
Cannula insertion depth: 4-7 mm (0.16-0.28 in)
Depth of insulin infusion: ≥ 4 mm (0.16 in)
Waterproof rating: IP28 (7.6 metres (25 feet) for up to 60 minutes)
Insulin concentration: U-100
Alarm type: Audible. Output: ≥ 45 db(A) at 1 metre
Operating relative humidity range: 20 to 85%, non-condensing
Storage relative humidity range: 20 to 85%, non-condensing
Operating atmospheric pressure: 700 hPA to 1,060 hPA
Storage atmospheric pressure: 700 hPA to 1,060 hPA
2 to 6 hours in 30-minute increments. Default is 4
hours.
0.05-30 U.
%, Units or Off. Default is Off.
30 minutes to 8 hours in 30-minute increments.
Maximum of 7. Cannot exceed the Maximum Bolus.
30 minutes to 2 hours.
10 to 50 units in 1-unit increments. Default is 10.0 U.
Off, or 1 to 24 hours in 1-hour increments. Default is
Off.
Rolling 90-day period.
English, German, French, Dutch, Italian, Finnish,
Danish, Swedish and Norwegian. Default is English.
Appendix
Effective Date: 27SEP2019, ECO-000240
177