3 Changing Your Pod

First Time Activating a Pod

Your nurse/educator will help you set up your first Pod.
After completing the PDM setup flow, the Home screen displays "No Active Pod".
1. Tap SET UP NEW POD.
2. Go to "Preliminary Pod Change Steps" on page 40.

Deactivate an Active Pod

To deactivate and remove an active Pod:
1. Navigate to the Pod change screen:
Home > Pod Info tab > VIEW POD
DETAILS
or
Menu icon (
2. Tap CHANGE POD, then tap
DEACTIVATE POD.
If a temp basal or extended bolus was
running, it will be cancelled now.
If you see a communication error
message, see "Error when deactivating
a Pod" on page 122.
Warning: Do NOT apply a new
Pod until you have deactivated
and removed the old Pod. A Pod
that has not been deactivated
properly can continue to deliver
insulin as programmed, putting
you at risk of over infusion and
possible hypoglycaemia.
38
) > Pod
Effective Date: 27SEP2019, ECO-000240