4. Probe Compensation 
When  connecting  the  probe  with  any  input  channel  for  the  first  time,  it  is  required  to 
make the adjustment, matching the probe with the input channel. Probe without being 
compensated and calibrated will cause measurement error or mistake. To adjust probe 
compensation, follow the steps below:   
1. Set the probe menu attenuation coefficient as 10×, place the switch on the probe at 
10×  and  connect  the  oscilloscope  probe  with  CH1.  Connect  the  probe  end  to  the 
signal output connector of the probe  compensator,  and connect  the ground clamp 
to the ground lead connector of probe compensator. Turn on CH1 and press AUTO.   
 
2.    Observe displayed waveform   
Over‐compensation 
 
 
 
If  the  screen  displays  "under‐compensation"  or  "over‐compensation"  for  waveform  as 
shown  in  above  figures,  use  screwdriver  with  nonmetal  handle  to  adjust  the  variable 
capacitance on the probe until screen displays "correct compensation" for waveform as 
shown in the below figure.   
 
 
 
 
Figure 1‐7 Probe Compensation Signal 
Correct compensation 
Figure 1‐8a Probe Compensation and Correction 