Approved for a body weight of up to 100 kg.
The product was developed for use in interim prostheses.
The product is suitable for use in a definitive TF prosthesis.
2.3 Environmental Conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service Life
In general, all modular adapters are tested by the manufacturer for 3 million
load cycles. Depending on the patient's level of activity, this corresponds to
a service life of 3 to 5 years.
3 Safety
3.1 Explanation of Warning Symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
3.2 General Safety Instructions
CAUTION
Excessive strain on the product
Fall due to breakage of load-bearing components
► Use the prosthetic components according to the classification (see
section "Area of Application").
CAUTION
Unallowable combination of prosthesis components
Fall due to breakage or deformation of the product
► Only combine the product with prosthesis components that are
approved according to the section "Combination Possibilities".