Product Description

1 Product Description
Notes regarding the document
see fig. 1 (see page 2)
INFORMATION
Last update: 2015-02-02
► Please read this document carefully before using the product.
► Follow the safety instructions.
► Please keep this document in a safe place.
These instructions for use are also available on the tablet in the product app (for further information please see the
section "Opening the product instructions for use in the app" - Opening the product instructions for use in the
app).
The use of the tablet is described in the corresponding user manual. The manual is available on the tablet in the
product app (for further information please see the section "Opening the tablet user manual in the app" -Opening
the tablet user manual in the app).
Relationship between the figure pages and the chapters
The figure pages at the beginning of this document contain figures about the product and its use. The figures are
related to the chapters.
For easier navigation, a chapter reference (chapter number below the symbol of the title page of the document) is
located below the figure number of each figure, and a figure reference (line "•   see Fig." with figure number) is
included in the chapters.
The figure reference is always located above the related text.
Proper use
see fig. 2 (see page 2)
The 3D L.A.S.A.R. Posture is a measuring device intended for use in the fields of prosthetics and orthotics.
The bench alignment of the prosthesis or orthosis is carried out in the workshop. The prosthesis or orthosis is
placed on the patient in the fitting room. The patient stands on the force sensing platforms for measuring. The
measured values are evaluated by a mini-computer in the lower part of the case and their shape superimposed on
the images from the cameras in the stands in the form of load lines. Depending on the settings, the load lines are
based on either vertical ground reaction forces alone, or on a combination of vertical and horizontal ground reac­
tion forces.
The images with the load lines are displayed in the app in the tablet. Measurement lines can also be determined
parallel to the load lines in order to test the alignment recommendation values and, if necessary, to implement and
check the changes necessary to optimise the static alignment of the prosthesis or orthosis.
Images can be saved from the monitor view and assigned to a patient in the app's patient database.
40 | Ottobock