Wiring Schematic #1 - Xylem MJK Oxix Installation And User Manual

Dissolved oxygen transmitter
Table of Contents

Advertisement

Dissolved Oxygen Transmitter

Wiring Schematic #1

for " Magflux
with Blind Lid and Oxix
®
page (page 32).
Oxix
with Display Unit
®
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot
"Display -"
Magflux
with Blind Lid
®
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot
"Display -"
GB Oxix Manual 1209
with Display Unit" from the opposite
®
33
GND A
B
+DC
Sensor supply
Fuse
200 mA T
+
A
B
-
+
-
C
-Display-
-mA-
-DI-
1
2 Shield
1
2 Shield
FW: 841013-009, 846001-002, 841513-009
Max. four MJK converter
can be connected to one
Converter #1
Conv
2 wire
shielded cable.
NO
C1
N01
C2 N02
-D01-
-D02-