Installation Of Cable Duct; Modification Of Cable Outlet Cover - Siemens Opdima MAMMOMAT 1000 Installation And Start-Up Instructions Manual

Table of Contents

Advertisement

5 - 16

Installation of cable duct

Four cables are laid from the bottom of the biopsy controller in a cable duct to the cable
outlet cover (3/Fig. 17). To install the cable duct on the rear side of the stand proceed as
follows:
1.
Fig. 17 Cable duct

Modification of cable outlet cover

To allow the cables to pass through the cable outlet cover, the cover has to be modified
according to the following:
1.
2.
3.
4.
Opdima?
Installation
Attach the cable duct (2/Fig. 17) to the rear cover by removing the protective strips
from the attached Velcro strips and place the duct flush with the biopsy controller
(1/Fig. 17). Secure the cable duct to the two standoffs (5/Fig. 17) with the
delivered screws and washers.
4
Cut an opening centrally on the cable outlet cover (4/Fig. 17). The opening should
be approximately 50 x 25 mm.
Attach the supplied plastic strip around the opening of the cable outlet cover
(4/Fig. 17).
Install the cable outlet cover (4/Fig. 17).
Mount the previously removed stand covers.
Register 3
5
SPB7-230.033.10
Rev. 03
08.02
Installation
UP
Page 16 of 26
SPS-UD
5
1
2
3
5
Siemens-Elema AB
Solna, Sweden