Braun VISTA Series Installation & Service Manual page 7

Commercial wheelchair lift
Hide thumbs Also See for VISTA Series:
Table of Contents

Advertisement

The VL917 can be configured for use with a cus-
tomer supplied remote interlock with either a +12V
signal (positive logic) or a ground signal (negative
logic). The lift is shipped ready to use in the nega-
a.
GND
J4
Positive
J6
Negative
J13
Red
J14
White
J2
J5
Circuit Board
(When installing a remote interlock that supplies a +12V signal)
1. To connect remote interlock switch:
a. Relocate male plug 3-wire jumper from J6
(Negative) to J4 (Positive) position on circuit
board located inside pump module cover.
b. Install a female insulated spade terminal on
end of customer supplied remote interlock
+12V source wire. See Specifications below.
c. Mounted to the pump module cover is a female
interlock plug. Exiting the back side (inside
pump cover) is a black and red wire with 1/4"
insulated male spade terminals.
d. Plug the +12V source from remote interlock
12V
BATTERY
-
+
COM
POSITIVE LOGIC
INTERLOCK SWITCH
• 15 AMP rating
• Customer supplied
Series 02 VL917IB Supplement 29682
Series
02
VL917IB
Series
02
VL917IB
Lift Interlock Installation
12V
J1
To
Remote
Interlock
J3
Positive Logic (+12) Interlock Installation Instructions
Specifications
Customer
Supplied Wire*
* Use 16 gauge wire
NC
(Red preferred)
NO
Supplement 29682
tive logic mode. To convert to a positive logic
interlock mode, follow the instructions below. See
supplement page 4 for negative logic mode hookup.
b.
d.
To Circuit
Board
into red wire supplied with insulated male
spade connector.
e. Remove male interlock jumper from front of
pump module cover. Store in glovebox for use
if needed when troubleshooting interlock
circuit. Note: Installing male interlock jumper
(e) will bypass customer-installed interlock
circuit and allow lift operation.
2. With lift unoccupied, test all lift functions to ensure
proper interlock installation (with interlock switch
opened and closed).
Customer
Supplied
Terminal
RD
To Circuit
Board
BK
Male
RD
Interlock
Jumper
RD
BK
c.
BK
Connect to 1/4" male
insulated spade
terminal on Red wire.
Interlock
Jumper
Page 5
e.