Thermo Scientific Lindberg/Blue M Mini-Mite TF55030 Installation And Operating Manual page 45

Tube furnace
Table of Contents

Advertisement

| Main Controller: 1x8 & 5x16 Segment Programmable
Chapter 6
|
6-43
304243H02
3. Press the UP
or DOWN
edit an existing program. The scrolling display shows "CURRENT PROGRAM
NUMBER".
End Type parameter
The action which occurs at the end of program or in reset depends on the configuration of
the 'END.T' parameter. The 'END.T' can be:
dwEll : Controls at last program setpoint.
SP2 : Controls at setpoint 2 (When the programmer completes the target setpoint will
switch to setpoint 2. The setpoint 2 may be a lower or a higher temperature).
rES : Reset on completion and reverts to SP1 or SP2, based on the Setpoint selection.
Scroll
through parameters in Level 2 and set the required 'END.T' by pressing
or
.
OR
button to select a number for a new program or to
OR
Mini-mite Tube Furnace

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lindberg/blue m mini-mite tf55035

Table of Contents