| Main Controller: 1x8 & 5x16 Segment Programmable
Chapter 6
No
Parameter
.
1
T.STAT
2
T.REMN
3
T.ELAP
4
A1.DHI
5
A2.HI
6
SP.SEL
7
SP1
8
SP2
9
RAMPU
10
SP.RAT
11
A.TUNE
12
UNITS
13
TM.CFG
14
SS.SP
15
SS.PWR
16
T.T
17
THRES
18
END.T
19
Timer.TimeRes
20
21
TSP.1 to TSP.4
22
RMP.1 to RMP.4
23
DWEL.1 to DWEL.4 Dwell Time 1 to Dwell Time 4
24
PB
25
TI
26
TD
27
LBT
28
PV.OFS
|
6-33
304243H02
The 8-segment programmable 3216c controller consists of microprocessor based
three-mode PID (Proportional, Integral and Derivative) and appropriate output switching
devices to control the furnace. The programmable controller can be used as a single set
point controller or as a programmable controller. The 8-segment digital model enables
eight segments of programming. The controller is capable of varying temperature or
process value with time through programming. A program is stored as a series of
segments and can be run once, repeated a set number of times or run continuously.
Some of the commonly used parameters in operator Level 1 and Level 2 of 3216c
controller are:
Table 2 Parameter Description and Accessibility in 3216c
Description
Timer Status
Timer Remaining
Elapsed Time
Deviation High Alarm Set Point
High Temperature Alarm Set Point
Set point Select
Set point 1
Set point 2
Set point Ramp Units
Set point Rate Limit
Auto Tune Enable
Display Units
Timer Configuration
Soft Start Setpoint
Soft Start Power Limit
Requested Time Duration
Timer Start Threshold
Timer End Type
Timer resolution
Target Set point 1 to Target Set point
4
Ramp Rate 1 to Ramp Rate 4
Proportional Band
Integral Time
Derivative Time
Loop Break Time
PV Offset
Level
Access
Level 1 + 2
Read/Write
Level 1 + 2* Read Only*
Level 1 + 2* Read Only*
Level 1
Read/Write
Level 2
Read Only
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1 + 2* Read Only*
Level 1 + 2* Read Only*
Level 1 + 2* Read Only*
Level 1 + 2
Read/Write
Level 1 + 2
Read/Write
Level 1
Read Only
Level 2
Read/Write
Level 2
Read/Write
Level 2
Read/Write
Level 2
Read/Write
Level 2
Read/Write
Level 2
Read/Write
Level 2
Read/Write
Level 2
Read/Write
Level 2
Read/Write
Page
Value
No.
rES
6-31
-
6-32
-
6-32
50
6-30
1125
6-30
SP1
6-23
0-1100
6-23
0-1100
6-23
Mins
6-23
Off
6-24
Off
6-25
Deg C
6-24
Prog(10)
6-37
-
6-34
-
6-34
-
6-35
1
6-36
Dwell
6-37
6-37
Mins
6-37
550
6-43
OFF
6-43
1 hour
6-44
15
-
95
-
16
-
30 mins
6-30
0
6-29
Mini-mite Tube Furnace