Mini-mite Tube Furnace
Maintenance & Cleaning
CAUTION : Maintenance should only be performed by trained personnel.
WARNING : Disconnect furnace from main power before attempting any
maintenance to furnace or its controls.
WARNING : Before maintaining this equipment, read the applicable SDS (Safety
Data Sheets). SDS is provided with the unit.
WARNING : When installing, maintaining or removing the refractory insulation, the
following precautions will minimize airborne dust and ceramic fiber:
Keep personnel not involved in maintenance out of the area.
Use a good vacuum to clean area and equipment. Do not use compressed air.
Use NIOSH high efficiently respirator (3M #8710 or equivalent).
Wear long sleeve clothing, gloves, hat and eye protection to minimize skin and eye
contact. Do not wear contact lenses.
Thoroughly wash self after work is complete.
Launder work clothing separate from other clothes and thoroughly clean laundering
equipment after use. If clothing contains a large amount of dust and/or ceramic
fiber, dispose of rather than clean.
Promptly place used ceramic fiber parts and dust in plastic bags and dispose of
properly.

Thermocouple Replacement

WARNING : Disconnect furnace from main power before attempting any
maintenance to furnace or its controls.
NOTE For optimal performance, the thermocouple should be replaced once a year. In
some situations a more frequent replacement schedule is recommended. S.br on the
controller display indicates a broken thermocouple.
Refer to Figure 3 "Thermocouple Replacement" as you perform the following procedure:
1. Disconnect the main power.
2. Open the latch.
Maintenance & Cleaning |
Chapter 8
304243H02 | 8-48