4
Operation
4.1
Function check
Perform a complete function check of the tapping unit prior to use.
• The tube connectors are firmly secured and no mechanical forces are acting on the tubes.
• The plastic and rubber product parts are in perfect condition and show no signs of ageing.
• The accessories are correctly connected.
• The product is leak-free.
• It is possible to adjust the flow rate from zero to the maximum flow.
• The product is in perfect hygienic condition.
4.2
Setting the flow for treatment
DANGER!
Function check!
The product is used in the treatment of patients. Any restriction in the unit's
performance can result in serious complications in treatment.
Perform a complete function check every time before using the unit.
WARNING!
Compressed gas setting!
The regulating mechanism is sensitive. Make the compressed gas settings very
carefully!
NOTE
When applying oxygen in its function as a medication as per the monograph in the
European Pharmaceuticals Reference, it is absolutely necessary to measure the
flow rate.
NOTE
The flow rate can be set to the following 14 values: 0; 0.2; 0.5; 0.7; 1; 1.5; 2; 3; 4;
5; 7; 10; 12; 15.
4
Operation
Function check
15