7

Maintenance

General

7
Maintenance
7.1
General
Maintenance, repairs and periodic tests may only be carried out by persons who have the
appropriate technical knowledge and are familiar with the product. To carry out these measures,
the person must have the necessary test devices and original spare parts.
ATMOS recommends: Work should be carried out by an authorised ATMOS service partner. This
ensures that repairs and testing are carried out professionally, original spare parts are used and
warranty claims remain unaffected.
7.2

Periodic tests

At least every 5 years a test must be performed.
7.3

Malfunctions and troubleshooting

Defect
No flow
Flow does not stop
The value of 15 l/min cannot
be achieved even if the flow
rate is adjusted to the max.
value.
Full loss of pressure at the
flowmeter
Tab. 5:
22
GA 5752 5678 GB 07
DANGER!
Health hazard!
The product is used in the treatment of patients. The product or some of its
components may be contaminated.
Clean and disinfect the product before maintenance and repair.
Repair work may be performed by personnel authorised by ATMOS.
Source of malfunction
Flow rate is set to the value
'0'
Product defective
Flow rate is not set to the
value '0'
Product defective
The pressure from the central
gas supply system is not high
enough
Sintered filter is clogged
Seal is missing or defective
There is a leak in the housing
Accessories are not tightened Check fit of accessories
Troubleshooting
Troubleshooting
Set the flow rate to the desired
value
Have the equipment repaired by
a service technician authorised
by ATMOS
Set the flow rate to the value '0'
Have the equipment repaired by
a service technician authorised
by ATMOS
Check the pressure supplied by
the central gas supply installation
Have the equipment repaired by
a service technician authorised
by ATMOS
Have the equipment repaired by
a service technician authorised
by ATMOS