Chair M 2 and are therefore a must besides regular cleaning. ® Repair work and period tests may be carried out only by expert personnel authorised by ATMOS. By applying only original spare parts you will have the guarantee that operational safety, readiness for work and the value of your ATMOS Chair M 2 will be preserved.
Graphic symbols contained in these operating instructions Warning, special diligent notice ) Important information Symbols of ATMOS Chair M 2 ® Degree of protection type B Order number This product complies with the relevant requirements of EU Directive...
1.5 Scope of supply • Prior to dispatch, the ATMOS Chair M 2 was subjected to an extensive functional test and was carefully packed. ® Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note).
• The user must be familiar with the operation of the chair. corresponds to the value of your local mains supply. • ATMOS is not liable for personal injury and damage to • For mains supply, only use the power cable supplied (or an property if equivalent one).
3.0 Setting up and starting up 3.1 Assembly • Always place the device on level, solid surface. 3.2 Starting up • Set up product at the desired spot. Any floor unevenness must be compensated for. • Check that the upper part of the chair can rotate freely. •...
3.0 Setting up and starting up 3.3 Electrical connection The ATMOS Chair M 2 is supplied with a power cable and ® IEC connection. Insert the power cord to the IEC socket at the back side of the chair base and the wall mounted power outlet.
= Up = Down Furthermore, the ATMOS Chair M 2 features an “Auto – ® Down (homing)” function, to lower the upper part of the chair down to its lowest level after a short tap on the foot switch.
4.0 Operation 4.5 Adjusting the headrest • Loosen the arresting button on the rear of the backrest by turning it anti-clockwise, • Adjust the headrest to the desired height (it may not be extended by more than 10 cm), • Arrest the headrest by turning the arresting button clockwise.
• The surfaces of the ATMOS Chair M 2 are resistant against all the recommended surface disinfectants stated in chapter ® 5.3 and 5.4. Nevertheless after any length of time discolourations could possibly develop. Polar solvents (e.g. acetone or chlorinated hydrocarbons (CCs)) may not be used for cleaning and disinfecting.
5.0 Cleaning and care 5.3 Recommended surface disinfectants • The surfaces of the ATMOS Chair M 2 can be cleaned / wiped with disinfectants containing the following active ingredients: ® - QAV (quarternary ammonium compounds) • Substances such as blood need to be removed immediately to prevent stains.
• Check connection at wall socket • No mains voltage • Check main fuse • Check mains plug on perfect fit at the device • Defect fuse • Exchange of fuse In case of malfunctions, please contact your local authorized ATMOS service technician.
9.0 Technical data Voltage 230 V~ ± 10 %; 50 Hz Special voltage 120 V~ ± 10 %; 60 Hz Current consumption Max. 2.5 A Power consumption Max. 520 VA Fuses T 3.15 A / 250 V Operating time 1.5 min operation, 8.5 min rest period Dimensions H x W x D 127.5 x 63.0 x 84.0 cm Weight...
10.0 Disposal • The ATMOS Chair M 2 does not contain any hazardous substances. ® • The housing is recyclable. • Device and accessories must be decontaminated prior to disposal. • Pay attention to a careful separation of the different materials.
11.1 Guidelines and Manufacturer´s Declaration - Emissions The ATMOS Chair M 2 is designed for operation in the environment specified below. The customer or user of the ATMOS ® ® Chair M 2 should ensure that it is used in such an environment.
Page 18
UT is the alternating mains voltage prior to application of the test levels. 11.3 Guidelines and Manufacturer's Declaration - Immunity The ATMOS Chair M 2 is designed for operation in the electromagnetic environment specified below. The customer or user of ® the ATMOS Chair M 2 should ensure that it is used in such an environment.
Page 19
Chair M 2 ® The ATMOS Chair M 2 is intended for use in an electromagnetic environment in which HF disturbances are controlled. The ® customer or user of the ATMOS Chair M 2 can thereby help to prevent electromagnetic interference by maintaining a minimum ®...