3.1 Scope of supply 3.2 Connections 3.3 Mounting the holder for light handle 3.4 Startup 3.4.1 Connection diagram ATMOS Cam 21 / 31 DATA and ATMOS Strobo 21 LED Operation................12 4.1 Functions of display 4.1.1/2 Display during and after phonation 4.1.3 Adjustment of display...
1.1 Notes on operating instructions In the following all device versions, as well as all modules and table-top devices are named ATMOS Strobo 21 LED! These operating instructions contain important notes on how to operate the ATMOS Strobo 21 LED safely, correctly and effectively.
...1.0 Introduction 1.4 Explanation of symbols Short cuts / symbols contained in this manual Move, plug ... Follow the arrows whilst in this direction proceeding, sequence Turn, shift ... Please press where in this direction the dot indicates Activate the optional foot switch Replace Check...
For your safety ● The ATMOS Strobo 21 LED is: The ATMOS Strobo 21 LED may only be used under the ● designed in line with IEC 601 / supervision of skilled staff who have been authorised EN 60601. by ATMOS and trained in its operation (IEC 601-1 / EN assigned to VDE safety class I 60601-1).
■ Airborne sound microphone incl. clip for ■ Holder for laryngoscope (1 quiver) laryngoscope ■ Light handle ATMOS LS 21 LED incl. adapter Wolf ■ Holder for light handle LS 21 LED with switching function (not applicable for ATMOS Roadster) ■...
Strobo 21 LED Cam 21 / 31 Cam 31 DV Data When applying the ATMOS Cam 21 / 31 for video stroboscopy please pay attention to connecting the dongle connector (Ref 507.4781.0) with the interlink connection of ATMOS Cam 21 / 31 before...
Operation 4.1 Functions of display 4.1.1 Indication during examination 4.1.2 Indication after examination Frequency Ø-value (pitch) mean frequency and Ø-volume after phonation Volume (Sound pressure level) Change of contrast ...or... push repeatedly 4.1.3 Setting menu Access „Setting Menu“ Select menu Select release submenu...
... 4.0... Operation Adjustment of menu Verify / Back Changing to the relative the submenu submenu Display indicates: „Please switch off and on again“ Exit the menus, back to indication for examination 4.2 Examination 4.2.1 Operating sequence click click click read off Phonation measured...
In case the picture gets more blurred when encreasing the brightness: The ATMOS Strobo 21 LED fl ashes with 2 Watt. For generating a brighter image, the illumination has to be longer. This entails that the vocal cord oscillates fractionally at this moment. The tiny movement reclined during the exposure time is perceived blurred.
Cleaning 5.1 General information on cleaning and disinfection ■ Compositions of plastics, varnish and colours as well as cleanser and disinfectants are changing ongoing. Therefore prior to fi rst use please test all cleaning agents at the undersurface! Prior to cleaning ■...
Maintenance and servicing Medical devices must always offer a maximum in safety and function. We therefore recommend: Prior to each application: necessary 6.1 Replacing the fuse click click Always replace both fuses! . . . 6.2 Regular safety-related controls The following safety-related controls have to be executed at a minimum all 12 months from persons, who are able to execute these controls due to their apprenticeship, their knowledge and due to their gained practical experience.
Please also pay attention to the stated notes there about the topic „Resolution / Brightness control“. Dongle connector for interlink Plug in dongle connector to ATMOS Cam 21 / 31. ATMOS Cam 21 / 31 is not plugged in. Foot regulator does not work Foot regulator is not connected Check if foot regulator is connected properly.
Technical specifi cations ATMOS Strobo 21 LED Operating voltage 100 - 240V, 50/60Hz Power consumption (max.) 0,3A Input power 30 VA Flash frequency 70 - 1000 Hz Accuracy of displaying fl ash frequency +/- 1 Hz Metering range of sound level...
The materials of the housing can be recycled completely. ■ The ATMOS Strobo 21 LED does not contain any hazardous goods. ■ The component parts of the ATMOS Strobo 21 LED must be disposed of correctly and the materials are to be separated carefully.
11.1 Guidelines and Manufacturer´s Declaration - Emissions The ATMOS Strobo 21 LED is intended for use in the electromagnetic environment specifi ed below. The customer or user of the Strobo 21 LED should ensure that it is used in such an environment.
Page 22
11.3 Guidelines and Manufacturer´s Declaration - Immunity The ATMOS Strobo 21 LED is intended for use in the electromagnetic environment specified below. The customer or user of the Strobo 21 LED should ensure that it is used in such an environment.
Page 23
If the measured field strength at the location where the ATMOS Strobo 21 LED is used exceeds the above compliance level, the ATMOS Strobo 21 LED is to be observed to verify the intended use. If abnormal performance characteristics are noted, additional measures might be necessary, e. g. a changed arrangement or another location for the device.
Need help?
Do you have a question about the Strobo 21 LED and is the answer not in the manual?
Questions and answers