Table 7. Electromagnetic Immunity - Nonin Sensmart H500 Operator's Manual

Multi-sensing portable oximetry system
Table of Contents

Advertisement

Immunity Test
This device is intended for use in the electromagnetic environment specified below. The
customer and/or user of this device should ensure that it is used in such an environment.
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
NOTE: U
is the AC mains voltage before application of the test level.
T
40

Table 7. Electromagnetic Immunity

IEC 60601 Test
Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
±5% U
(>95% dip
T
in U
)
T
for 0.5 cycle
±40% U
(60% dip
T
in U
)
T
for 5 cycles
±70% U
(30% dip
T
in U
)
T
for 25 cycles
<5% U
(>95% dip
T
in U
)
T
for 5 seconds
3 A/m
Compliance
Level
Environment—Guidance
±6 kV
Floors should be wood, concrete,
contact
or ceramic tile. If floors are
covered with synthetic material,
±8 kV air
relative humidity should be at
least 30%.
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Technical Information