System Operation

System Operation
CAUTION: Between patients, turn the H500 system to Sleep Mode.
!
When the device is turned on, the monitor begins a new patient record.
WARNING: This device is intended only as an adjunct device in patient
assessment. It should not be used as the sole basis for diagnosis or
therapy decisions. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
WARNING: As with all medical equipment, carefully route patient cables
and connections to reduce the possibility of entanglement, strangulation,
or tripping.
WARNING: Display and signal processor cannot be used outside of
carrying case.

Initial System Set-Up

Follow the below steps when setting up the system for the first time.
1.
Fully charge display and signal processor per instructions in "Charging
the System" section on page 23.
2.
Press and hold the On/Off, Sleep button on signal processor until red,
yellow, green LEDs blink on and Blue Bluetooth LED begins to blink.
3.
Press and hold the On/Off, Sleep button on display until screen turns on.
4.
It is recommended to set Date and Time following "Setting Date and
Time" instructions in "Setting Date and Time" section on page 29 upon
initial setup of device.
5.
Connect sensor to signal processor and engage sensor lock.
6.
Observe display during start-up to ensure screen is working and
expected parameters are indicated.
7.
System is now ready to take a measurement or to be turned off for future
use.

Start-up Sequence

Each time the monitor is turned on, it performs a brief start-up sequence.
1.
Press On/Off, Sleep button on display.
2.
The display lights up and displays the welcome screen.
3.
Sensor and signal processor turn on.