Table 10. Electromagnetic Immunity - Nonin Avant 9700 Operator's Manual

Digital pulse oximeter
Hide thumbs Also See for Avant 9700:
Table of Contents

Advertisement

Technical Information
Immunity Test
This device is intended for use in the electromagnetic environment specified below. The customer and/or
user of this device should ensure that it is used in such an environment.
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power Frequency (50/
60 Hz) Magnetic Field
IEC 61000-4-8
NOTE: U
is the AC mains voltage before application of the test level.
T
Table 10: Electromagnetic Immunity
IEC 60601
Test Level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
±5% U
(>95% dip in
T
U
) for 0.5 cycle
T
±40% U
(60% dip in
T
U
) for 5 cycles
T
±70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in
T
U
) for 5 sec.
T
3 A/m
Compliance Level
±6 kV contact
±8 kV air
±2 kV
±1 kV
±1 kV
±2 kV
±5% U
(>95% dip in
T
U
) for 0.5 cycle
T
±40% U
(60% dip in
T
U
) for 5 cycles
T
±70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in U
)
T
T
for 5 sec.
3 A/m
Electromagnetic
Environment—Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, relative
humidity should be at least
30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.