13 Help messages
Code
H161
Temperature
H162
H165
H166
H170
User related
messages
H171
H181
H182
H186
H187
H189
H190
H191
H192
H193
H194
H195
H196
H197
H199
68
Planmeca Viso G7
Explanation
The temperature of the tube head is
too high.
The temperature of the lift motor is
too high.
The temperature of the tube head is
too high for the selected exposure
values.
The maximum tube head energy
level was exceeded.
Wrong licence code.
Timeout in production test.
The imaging process was cancelled
in Planmeca Romexis.
Timeout in image data transmission. Exposure was interrupted.
No IP address defined for 3D
sensor.
Problem during image data
transmission.
The screen was touched during
exposure.
Protouch-CPU communication
failure.
3D sensor communication failure.
Workstation communication failure.
Invalid scan settings.
CPU connection not established.
Request timed out while waiting for
CPU to respond.
Version mismatch in communication
interfaces.
Workstation communication
disabled.
Video streaming failed.
Comments
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the
lift motor to cool down.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the
tube head to cool down or
use lower exposure values.
Check the licence code.
Contact your service
technician for help.
Exposure was interrupted.
Contact your service
technician for help.
Exposure was interrupted.