Linearity of radiation output
Exposure time
3D
Pan (SmartPan)
ProCeph
SID
3D / Pan (SmartPan)
Ceph
Magnification
3D
Pan (SmartPan)
Ceph
Duty cycle for height adjustment
Line voltage
Line current
Input power
Stand by
Exposure
Line harmonics
Max. permissible apparent impedance of supply
mains
Max. continuous heat dissipation
Internal fuse(s)
One user replaceable fuse
Type
External fuse(s)
User's manual
18 Technical specifications
Dose range min. / max. as indicated on system
user interface.
Accuracy of dosimetric indication (DAP, CTDI):
±40%
< 0.1
Pulsed, effective 1.5 ‐ 36 s as indicated ±10%
2.5 – 15.6 s as indicated ±10%
0.1 – 1.6 s as indicated ±10%
700 mm
1700 mm (66.9 in.)
1.40 ‐ 1.71
1.40
1.13
25 s ON / 400 s OFF
100 ‐ 220 V~ / 50 ‐ 60 Hz
230 ‐ 240 V~ / 50 Hz
8 ‐ 17 A
150 VA
1800 W
Cos better than 0.9
0.5 ohm (100 VAC)
250 W
100 ‐ 220 V~ / 16A FF H 500 V
230 ‐ 240 V~ / 8A FF H 500 V
195100 ELU
100 ‐ 220 V ~ / 16A min. - 20A max. T 250 V
230 ‐ 240 V ~ / 10A min. - 20A max. T 250 V