WRITE TO UNIT

After you have entered configurations,
selected and configured the Tank Shape and
entered the Tank Values, click "Write to Unit"
and load the configuration into the memory of
the sensor. When completed, this
configuration will remain inside the sensor
memory and will not change unless the sensor
is connected to WebCal
configuration is written to the sensor. Loss of
power will not change or lose the configuration
within sensor memory.
Next, use the file management features to save your configuration by clicking "Save Config File" and print
your wiring diagram by clicking "Wiring Diagram."
"Save Config File" will save this configuration as a text file which can be loaded back into WebCal
pressing the "Open Config File" button. It is good practice to save the configuration file for each different
configuration with a unique name for easy identification. If using multiple sensors in identical applications, then
use of a single configuration file is recommended.
"Wiring Diagram" will display a PDF file showing the unique wiring for the specific configuration created in
®
WebCal
. The PDF can be printed or emailed. It is good practice to save the wiring diagram as a backup
"Factory Reset" will return the sensor to its original factory configuration. The factory probe lengths will return
to the original maximum range. Sensors with pre-cut lengths of 36" (91.4cm), 72" (182.9cm) and custom
lengths will also return to the original factory configuration of 118.1" (300cm) for Rod and Coaxial probes and
216.5" (549.9cm) for Cable probes. Note: Make sure to re-enter the correct probe length and click on Write to
Unit before installing the EchoWave
Probe Style
Rod
Coaxial
Cable (includes
counter-weight)
|
42
®
and a new
®
.
Sensor
LG10-0_ _ _-_ _-_ _ _
LG10-1_ _ _-_ _-_ _ _
LG11-2_ _ _-_ _-_ _ _
Probe Length
118.1"
(300cm)
118.1"
(300cm)
216.5"
(549.9cm)