Tank Level Configuration; Write To Unit - FlowLine EchoWave LG11 Series Manual

Guided wave liquid level transmitter
Hide thumbs Also See for EchoWave LG11 Series:
Table of Contents

Advertisement

Configuration
C. Tank Level Confirmation
Verify the Height Units, Sensor Height, Probe Length, Maximum Fill-Height, Minimum Fill-Height, Capacity as
well as the Max./Min. Volume and Max./Min. Current. All values were calculated in the previous Dimensional
Entry window. To adjust these settings, click on Volumetric Modes.
D. Write to Unit
®
This WebCal
* operation uploads configuration into the sensor. Other features in the section include providing
a custom wiring diagram specific to the signal output and saving the configuration file to your hard drive.
Volumetric Sensor Output
The Volumetric Mode button will
be highlighted in Blue when a
volume output is selected.
* For complete information on the WebCal
flowline.com/webcal-software.
MN300840 REV A5
®
software, please refer to the WebCal
Section Five
®
manual located at
29
|