Preparation For Use; Initial Inspection - HP 8566B Installation And Verification Manual

Spectrum analyzer, includes option 400, option 462, and option 857
Hide thumbs Also See for 8566B:
Table of Contents

Advertisement

Warning
Warning
Warning
Caution
Preparation for
Use

Initial Inspection

2-2 Installation
Failure to ground the instrument properly can result in personal
injury. Before turning on the HP 8566B, you must connect its
protective earth terminals to the protective conductor of the
main power cable. Insert the main power cable plug only into a
socket outlet that has a protective earth contact. DO NOT defeat
the earth-grounding protection by using an extension cable,
power cable, or autotransformer without a protective ground
conductor. If you are using an autotransformer, make sure its
common terminal is connected to the protective earth contact of
the power source outlet socket.
Power is still applied to this instrument with the ILINE) switch
in STANDBY. There is no off position for the (LINE] switch. 'Ib
remove power from the instrument, it is necessary to remove the
power cable from the rear of each of the instrument sections.
Because the combined weight of the instrument sections is
approximately 112 pounds, use appropriate caution when moving
or installing.
Before switching on this instrument, make sure it is adapted to
the voltage of the ac power source as described in the following
procedures. Failure to set the ac power input to the correct voltage
could cause damage to the instrument when the ac power cable is
plugged in.
Inspect the shipping container for damage. If the shipping container
or cushioning material is damaged, keep it until you have verified
that the contents are complete and you have tested the instrument
mechanically and electrically.
Contents of the shipment are shown in Figure l-l. If the contents
are incomplete or if the instrument does not pass the operation
verification tests (procedures are provided in Chapter 5), notify the
nearest Hewlett-Packard office. If the shipping container is damaged
or the cushioning material shows signs of stress, also notify the
carrier. Keep the shipping materials for the carrier's inspection. The
HP office will arrange for repair or replacement without waiting for a
claim settlement.
If the shipping container and cushioning material are in good
condition, retain them for possible future use. You may wish to ship
the instrument to another location or to return it to Hewlett-Packard
for service. Chapter 1 provides instructions for repackaging and
shipping the instrument.

Advertisement

Table of Contents
loading

Table of Contents