General
HP-IB Interface
Functions
Environmental
Power Requirements
Humidity
X-Rays
Temperature
Operation:
0°C to 55°C
Storage:
-40°C to 75°C
Increased internal temperatures may result if the rear-panel air filters
are not cleaned regularly.
Altitude
Operation:
Storage:
515,240 m (50,000 feet)
50 to 60 Hz; 100, 120, 220, or 240 volts (+5%, -10%); approximately
650 VA (40 VA in standby). 400 Hz operation is available as Option
400.
Operation
Except as noted in electrical specifications, type tested at <95%
relative humidity, 25°C to 40°C for five days.
Storage
5% to 90% relative humidity, 0°C to 40°C.
Conducted and radiated interference is within the requirements
of MIL-STD 461C, Part 7 RE02 and CEO3 (Air Force), and
within the requirements of CISPR Publication 11 and
Messempfaenger-Postverfuegung 526/527/79 (Kennzeichnung Mit
F-Nummer/Funkschutzzeichen).
Serial Prefix 3004A and Above
X-rays generated by this instrument are sufficiently screened.
Die in diesem geraet entstehende roentgenstrahlung ist ausreichend
abgeschirmt .
accel. voltage / beschl. spg < 20 kV
Serial Prefix 3001A and Below
When operating, this instrument emits x-rays; however, it is well
shielded and meets safety and health requirements of various
countries, such as the X-Ray Radiation Act of Germany. Radiation
emitted by this instrument is less than 0.5 mR/hr at a distance of five
(5) centimeters from the surface of the cathode-ray tube. The x-ray
Specifications 3-17