4. Scale Fidelity
(Linear)
Related Performance
Description
Scale Fidelity Test
Test
A signal source of known amplitude is connected to the spectrum
analyzer and the analyzer is adjusted for a reference. The signal
source is stepped down from -10 dB to -30 dB in 10 dB steps, and
the amplitude of the displayed signal is measured using the marker
function. This measured value is used to calculate the percent error
from the established reference level.
4. Scale Fidelity (Linear)
Operation Verification 5-13