Amplitude
Measurement Range
Displayed Values
Measurement range is the total amplitude range over which
the analyzer can measure signal responses. The low value is
determined by sensitivity (10 Hz resolution bandwidth and 0 dB input
attenuation), and the high value by damage level.
Amplitude Measurement Range
Tuned Frequency
Non-Preselected
100 Hz to 50 kHz
50
Preselected
2.0
5.8
12.5 GHz to 18.6 GHz
18.6 GHz to 22 GHz
Scale
Over a lo-division CRT vertical axis with reference level (0 dB) at the
top graticule line.
10 dB/div for 90 dB display from reference level.
Expanded from reference level:
5 dB/div for 50 dB display
2 dB/div for 20 dB display
1 dB/div for 10 dB display
10% of reference level/div when calibrated in
voltage.
Readout expandable to +60.0* dBm to -119.9 dBm
(-139.9 dBm for 51 kHz resolution bandwidth) using
Linear
7.07 V to 2.2 PV full scale. Readout expandable to 223.6 V*
to 2.2 PV (0.22 ,LJV for <l kHz resolution bandwidth) using
*Maximum total input power not to exceed +30 dBm damage level.
-95 dBm to +30 dBm
-112 dBm to +30 dBm
-134 dBm to +30 dBm
-119 dBm to +30 dBm
-114 dBm to +30 dBm
Calibration
Calibration
Reference Level Range
Range
Specifications 3-5