Note
2-10 Installation
9. Insert the correct fuse, and push the fuse-pull lever into its
normal right-hand position.
10. Bolt each instrument section into the rack from its rack mount
flange. Place the IF-Display section just above the RF section.
11. Connect cable W31 between IF-Display section 52 and RF section
J6. See Figure 2-7.
12. Connect cable W30 between IF-Display section Jl and RF section
Figure 2-7. Rear-Panel Cable Connections
13. Connect the ac line power cords.
14. If the HP 8566B will be remotely controlled, refer to "To Set the
HP-IB Address" in this chapter.
15. When you turn your instrument on for the first time, you should
perform the verification tests in Chapter 5. Refer to the HP 8566B
Operating and Programming Manual for instructions on using
the instrument.
Cable W15 is normally connected between the FREQ REFERENCE
EXT and INT BNC ports, providing the HP 8566B with its own internal
10 MHz frequency reference. W15 is removed when an external
frequency reference is used. Cables W37 and W38 are removed for
connection of the IF and VIDEO ports to
Adapter. If the instrument is used without the HP 85650A Quasi-Peak
Adapter, W37 and W38 must be connected for
operate.
w 3 7
w 3 1
J 6
J 7
the
HP 85650A Quasi-Peak
the
instrument to