Note
Equipment
Connections
Note
The equipment needed to perform the HP 8566AD3 Operation
Verification is listed in both 'Ihble 5-l and 'Ihble 5-3. Table 5-l includes
a test-by-test listing of the equipment needed, and 'Ihble 5-3 includes
a test equipment summary. To obtain a permanent record of the test
results an HP-IB printer is required. The HP 8566A/B CRT display
may be selected to display the test results if a permanent test record
is not desired.
A change of the results for each test is expected over a period of
time; therefore, Hewlett-Packard warrants only the specification range
and not the repeatability of the data for any given specification.
For a dual bus system, connect the test equipment to the computer
as shown in Figure 5-l. The instrument under test is connected
separately to the HP 98624A Interface card. For a single bus system,
connect the test equipment and instrument under test to the
computer as shown in Figure 5-2.
Turn the HP 8566A/B (LINE) switch on and allow a 2 hour warm-up.
Also, allow sufficient warm-up time for the other test equipment as
indicated in the individual operating and service manuals.
When connecting signals from the synthesized sweeper to the test
setup, it is necessary to use a microwave test cable with minimum
attenuation to 22 GHz.
Figure 5-1. Dual Bus (MTS) System Connection
B U S t o U n i t U n d e r T e s t (HP8566B)
S e l e c t C o d e o f 8 .
Operation Verification 5-5