Specifications
Introduction

Frequency

Measurement Range

Displayed Values

Note
Note
Note
Unless noted, all specifications are for AUTOCOUPLED FUNCTION
operation and are with the preselector tracking optimized using the
MARKER PRESELECTOR PEAK function. Where specifications are
subject to minimization
limits are given unless noted. Nominal values provide useful, but
nonwarranted, information about functional performance.
100 Hz to 22 GHz, dc coupled input
Frequency Reference Error and Accuracy
Aging rate
Temperature stability ~7
frequency reference error,
The term
is defined as: f [aging rate/day x number of days since calibration +
temperature stability].
When the analyzer is in STANDBY, the frequency reference
temperature is maintained at a steady state. Frequency accuracy is
then subject to the standard instrument warm-up period indicated in
"General" in this chapter.
Changes in line voltage, gravitational field, and other environmental
conditions will affect

Center Frequency

0 Hz to 22 GHz
with
the error-correction routine, corrected
Accuracy
x lo-' over 0°C to 55°C range (25°C reference)
when used later in this manual,
the
frequency reference accuracy.
Specifications 3-l