Edit Standard - To edit data for an existing calibration standard,
move the highlight bar to the desired calibration standard's name
and press EDIT. The Calibration Standard Edit Mode will then be
entered. This mode is described below. The data for an existing
calibration can be deleted by pressing DEL while the highlight bar
is on the name of the standard.
Add New Standard - To define a new calibration standard, move
the highlight bar to the "NEW STANDARD" option and press
ADD. The Calibration Standard Edit Mode will then be entered.
This mode is described below.
.

Calibration Standard Edit Mode

This mode allows data for a new or existing calibration standard to
be edited. You may specify the name for the calibration standard
whether the reflectance values for the standard will be specified in
10 or 20nm bandwidths, and the actual reflectance values for the
standard. The reflectances may either be entered by manually
editing the values or measuring the actual calibration standard.
The standard name is edited in the same manner as other name
fields on the Elrepho. You should make sure the standard name is
meaningful to identify it when the name is displayed. If it is the
wrong standard, you can select the appropriate standard using the
Current Calibration Standard Selection Mode.
To store the reflectance values by measuring, you must first
calibrate the instrument using an appropriate calibration standard,
which is already defined. Once you have calibrated, position the
standard on the sample holder and press the Measure Button. After
the measurement, the reflectance values will be displayed.
Calibration Setup Mode
34