If you press CIE you will access the Illuminant Checker Calibration
Mode, which allows manual entry of Illuminant Checker's CIE
Whiteness value. Once the correct CIE Whiteness
value for the Illuminant Checker is displayed, the UVCAL step
number which produces the displayed CIE Whiteness value can be
determined by pressing the MEASURE button. Once determined,
the Illuminant Checker's CIE Whiteness and UVCAL step number
can be saved. The Illuminant Checker Calibration Mode requires
calibration in a UVINC measurement mode.
The GANZ key will only be displayed for instrument models which
support Ganz/Griesser Whiteness. If this key is pressed, the
Ganz/Griesser Parameter Calibration Mode is entered. This mode
allows the measurement of between 4 and 12 Ganz/Griesser
Whiteness Standards. You will be prompted for each standard's
Ganz/Griesser Whiteness value. Once the correct Whiteness is
entered for a standard, press the Measure key to perform a
measurement using the current UVCAL step position.
Note that the Ganz/Griesser Parameter Calibration mode requires
the instrument be calibrated in a UVINC measurement condition.
Repeat this process for each standard. After the 4th standard has
been measured, one soft key will be labeled COMP. Press this key
to compute the Ganz/Griesser parameters at the current UVCAL
step position. After the parameters are computed, they are
displayed in the Whiteness/UVCAL Setup Mode. If the parameters
are acceptable they can be saved. If not acceptable, the UVCAL
step position must be changed and the Ganz/Griesser parameter
calibration performed again.
If a new position is determined for UVCAL, you must
recalibrate. This applies to both CIE and GANZ/GRIESSER.

Language Selection

Use this mode to select the language to use for both displays and
printouts.