Measurement Mode - Datacolor Elrepho 3000 Manual

Table of Contents

Advertisement

Elrepho 3000
UV420 (420nm cutoff filter), UV460 (460nm cutoff filter), and
UVCAL (specific UV calibrator step) positions.
Some of the UV filter positions may not be calibrated when either
the "Current Job" or "All Jobs" option is selected (when jobs are
defined).
Once the calibration procedure is started, prompt messages will be
displayed when any actions need to be performed; i.e., position
white standard or black trap, adjust mechanisms.
You can press the MENU key to exit the calibration procedure.
The following is an example of a prompt message you will see to
measure the black trap for XLAV mode. The X's indicate the name
of the white reference calibration standard to be measured. See
"Calibration Setup Mode".

Measurement Mode

This mode is used to measure samples for the current job. It is
entered by pressing the MEAS key or the Measure Button while in
the Display Mode.
The current job's procedure is used to determine how the sample
and the instrument should be set up for each measurement required.
Prompts will be displayed if any operator action is required. When
prompted to perform a measurement, press the red MEASURE
button located to the left of the sample viewer.
14
Datacolor International