Control-IQ technology limits basal rate to 3 units/hour in the event of a transmitter error. In order to receive more than 3 units/hour during a transmitter error,
turn Control-IQ technology off.
232
Explanation
What does it mean?
How will the System notify me?
Will the System re-notify me?
How should I respond?
The transmitter has failed and the CGM session has stopped.
1 vibration, then vibration/beep every 5 minutes.
No.
Tap MORE INFO. A screen notifying you that your CGM session has
stopped but insulin delivery continues will appear.