CHAPTER 31 • Control-IQ Technology Alerts
31.2 Out of Range Alert – Control-IQ Technology On
Screen
What will I see on the screen?
WARNING
Control-IQ technology can only adjust insulin delivery when your CGM is in range. If you go out of range during insulin adjustment, your basal insulin delivery will
revert to the basal rate settings in your active Personal Profile, limited to 3 units/hr. In order to receive more than 3 units/hour while the sensor is not
communicating with the pump, turn Control-IQ technology off.
NOTE
Out Range Alert and Automated Insulin Dosing: It is recommended that you keep the Out of Range Alert turned on and set to 20 minutes. If your pump and
CGM are not connected for 20 minutes, the automated insulin dosing feature will not work. The automated insulin dosing feature will begin working immediately
when the transmitter and pump are back within range.
282
Explanation
What does it mean?
How will the System notify me?
Will the System re-notify me?
How should I respond?
Control-IQ technology is turned on, but the transmitter and pump
are not communicating. The pump will not receive sensor glucose
readings. Control-IQ technology will continue to adjust basal rates
and deliver automatic correction boluses for the first 20 minutes
that the transmitter and pump are out of range. Control-IQ
technology will resume automated insulin dosing once the
transmitter and pump are back within range.
1 vibrate, then vibration/beep every 5 minutes until the transmitter
and pump are back in range.
Yes, if the transmitter and pump remain out of range.
Tap
to confirm and move the transmitter and pump closer
OK
together, or remove the obstruction between them.