CHAPTER 30 • Configure and Use Control-IQ Technology
30.1 Required Settings
Required Personal Profile Settings
In order to use Control-IQ™
technology, the following Personal
Profile settings must be configured. See
Chapter 5 Insulin Delivery Settings
instructions about setting these values.
•
Basal rate
•
Correction Factor
•
Carb Ratio
•
Target BG
•
Carbohydrates turned on in Bolus
Settings
Required Control-IQ Technology
Pump Settings
In addition to the required Personal
Profile settings, there are two values
specific to Control-IQ technology that
must be set. These are:
•
Weight
•
Total Daily Insulin
270
Recommended Control-IQ
Technology Pump Settings
Although Sleep can be started and
stopped manually, it is recommended
that you schedule sleep. This chapter
explains how to do both. The following
settings are required to schedule sleep:
for
•
Selected Days
•
Start Time
•
End Time
30.2 Set Control-IQ Technology
Weight
Control-IQ technology cannot be
turned on unless Weight is entered. The
Weight value may be updated when
you visit your healthcare provider.
1. From the Home screen, tap
OPTIONS.
2. Tap My Pump.
3. Tap Control-IQ.
The Control-IQ screen is displayed.
4. Tap Weight.
5. Tap Pounds or Kilograms to set
the unit of weight.
6. Tap
.
7. Enter the weight value on the
numeric keypad.
8. Tap
.
9. If you are done with the Control-IQ
settings, tap
.
The SETTING SAVED screen is
temporarily displayed.