31.5 Max Insulin Alert

Screen
What will I see on the screen?
Explanation
What does it mean?
How will the System notify me?
Will the System re-notify me?
How should I respond?
CHAPTER 31 • Control-IQ Technology Alerts
The pump has delivered the maximum allowable 2 hour insulin
amount based on your Total Daily Insulin setting. You see this alert
when Control-IQ technology has delivered 50% of your Total Daily
Insulin (through basal and/or bolus deliveries) over the previous
rolling 2 hour window, and detects this condition for 20 minutes in a
row. Control-IQ technology will suspend insulin delivery for a
minimum of 5 minutes, and then resume insulin delivery once the
condition is no longer detected.
2 vibrations, then 2 vibrations/beeps every 5 minutes until
acknowledged.
Yes, every 5 minutes until acknowledged.
Tap
.
OK