CHAPTER 25 • CGM Alerts and Errors

25.16 Out of Range Alert

Screen
What will I see on the screen?
WARNING
Control-IQ technology can only adjust insulin delivery when your CGM is in range. If you go out of range during insulin adjustment, your basal insulin delivery will
revert to the basal rate settings in your active Personal Profile, limited to 3 units/hr. In order to receive more than 3 units/hour while the sensor is not
communicating with the pump, turn Control-IQ technology off.
230
Explanation
What does it mean?
How will the System notify me?
Will the System re-notify me?
How should I respond?
The transmitter and pump are not communicating. The pump will
not receive sensor glucose readings, and the automated insulin
dosing feature is not able to predict sensor glucose levels or adjust
insulin delivery.
1 vibrate, then vibration/beep every 5 minutes until the transmitter
and pump are back in range.
Yes, if the transmitter and pump remain out of range.
Tap
to confirm and move the transmitter and pump closer
OK
together, or remove the obstruction between them.