CHAPTER 32 • Overview of Control-IQ Technology Clinical Studies
DCLP5: Comparison of CGM Values Between Control-IQ and SAP Users (N=101)
Characteristic
Average Sensor Glucose
(std dev)
Average % 70–180 mg/dL
(std dev)
Average % >180 mg/dL
(std dev)
Average % <70 mg/dL
(std dev)
Average % <54 mg/dL
(std dev)
304
Control-IQ
SAP
162 mg/dL
179 mg/dL
(18 mg/dL)
(26 mg/dL)
67%
55%
(10%)
(13%)
31%
43%
(10%)
(14%)
1.8%
2.1%
(1.38%)
(1.18%)
0.34%
0.38%
(0.35%)
(0.35%)
Difference Between Study Arm and
Control Arm
-17 mg/dL
+11%
-10%
-0.40%
-0.07%