25.19 Failed Sensor Error

Screen
What will I see on the screen?
WARNING
Control-IQ technology limits basal rate to 3 units/hour in the event of a failed sensor. In order to receive more than 3 units/hour when a sensor fails, turn
Control-IQ technology off.
Explanation
What does it mean?
How will the System notify me?
Will the System re-notify me?
How should I respond?
CHAPTER 25 • CGM Alerts and Errors
The sensor is not working properly and the CGM session has
stopped.
1 vibration, then vibration/beep every 5 minutes.
No.
Tap MORE INFO. A screen notifying you that your CGM session has
stopped but insulin delivery continues will appear.
Replace the sensor and begin a new CGM session.