Overview; T:slim X2 Pump Specifications - Tandem t:slim X2 User Manual

With control-iq technology
Hide thumbs Also See for t:slim X2:
Table of Contents

Advertisement

CHAPTER 33 • Technical Specifications

33.1 Overview

This section provides tables of technical specifications, performance characteristics, options, settings, and electromagnetic
compliance information for the t:slim X2™ pump. The specifications in this section meet the international standards set forth in IEC
60601-1 and IEC 60601-2-24.

33.2 t:slim X2 Pump Specifications

t:slim X2 Pump Specifications
Specification Type
Classification
Size
Weight (with full disposable)
Operating Conditions
Storage Conditions
Atmospheric Pressure
Moisture Protection
Cartridge Volume
Cannula Fill Amount
320
Specification Details
External PSU: Class II, Infusion Pump. Internally-powered equipment, Type BF applied part. The risk of ignition of
flammable anesthetics and explosive gases by the pump is remote. While this risk is remote, it is not
recommended to operate the t:slim X2 pump in the presence of flammable anesthetics or explosive gases.
3.13" x 2.0" x 0.6" (L x W x H) - (7.95 cm x 5.08 cm x 1.52 cm)
3.95 ounces (112 grams)
Temperature: 41°F (5°C) to 98.6°F (37°C)
Humidity: 20% to 90% RH non-condensing
Temperature: -4°F (-20°C) to 140°F (60°C)
Humidity: 20% to 90% RH non-condensing
-1,300 feet to 10,000 feet (-396 meters to 3,048 meters)
IPX7: Watertight to a depth of 3 feet (0.91 meters) for up to 30 minutes
3.0 mL or 300 units
0.1 to 1.0 units of insulin