pump to calculate and add a correction
bolus to any other bolus you request.
To accept the correction bolus
press
. A correction bolus is
calculated and will be added to any
food bolus you request on the
Bolus screen.
To decline the correction bolus,
press
. No correction bolus will
be added to any food bolus you
request on the Bolus screen.
Below Target
If your BG or sensor glucose value is
below your Target BG, the pump
presents you with the option for the
pump to calculate and subtract a
correction bolus from any other bolus
you request.
To accept the correction bolus
press
. A correction bolus is
calculated and will be subtracted
from any food bolus you request on
the Bolus screen.
To decline the correction bolus,
press
. No correction bolus will
be subtracted from any food bolus
you request on the Bolus screen.
Within Target
If your BG or sensor glucose value is
the same value as your Target BG, no
Correction Bolus screen is displayed.
CHAPTER 7 • Manual Bolus
BG Value Manual Entry
If your sensor glucose value was not
auto-populated on the Bolus screen
based on the conditions needed for
that feature, you will need to enter your
BG value into the pump manually
before advancing to the Correction
Bolus confirmation screens. The
conditions needed for the auto-
populate feature are:
Control-IQ technology is turned on
and available
A CGM session is active
A CGM value is present
A CGM trend arrow is available on
the CGM Home screen
NOTE
For more information about CGM trend
arrows and how to use them for treatment
decisions, see the CGM manufacturer's
user guide. You can also see
Section 24.3
Rate of Change
Arrows.
The Correction Bolus confirmation
screens are displayed, if appropriate,
after you manually enter your BG value
on the Bolus screen.