Model DMM7510 7½ Digit Multimeter User's Manual
Install the GPIB cards in your computer
Refer to the documentation from the GPIB controller vendor for information about installing the
GPIB controllers.
Connect GPIB cables to your instrument
To connect a DMM7510 to the GPIB interface, use a cable equipped with standard GPIB connectors,
as shown below.
To allow many parallel connections to one instrument, stack the connectors. Each connector has two
screws on it to ensure that connections remain secure. The figure below shows a typical connection
diagram for a test system with multiple instruments.
To avoid possible mechanical damage, stack no more than three connectors on any one
instrument. To minimize interference caused by electromagnetic radiation, use only shielded
GPIB cables. Contact Keithley Instruments for shielded cables.
DMM7510-900-01 Rev. C / September 2019
Figure 14: GPIB connector
Section 3: Using a remote interface
3-3