Section 10: Capturing and analyzing waveforms

Power-up behavior

You can use the DMM7510 to capture the start-up behavior of the buck converter. The start-up
inspection can ensure that the device turns on in a reasonable amount of time without any
unexpected glitches or pulses. You can capture the asynchronous power-up event by using the
analog trigger feature on the DMM7510.
For this test, you:
Connect the 4 Ω load resistor to the output terminals of the buck converter
Connect test leads to the output terminals of the buck converter
Reset the instrument
Select the digitize voltage function and set the sample rate to 1 megasample per second
Set sample count to 10,000 to monitor the power-up event for 10 ms
On the front panel of the instrument, open the Graph screen and set the graph scale option
X-Axis Method to All
Set up the edge trigger level and position
Run the trigger model to capture the voltage waveform
Turn on 12 V of input voltage to the input terminals of the buck converter
Device connections
You can use either the front or rear inputs. Ensure that the front-panel TERMINALS switch is set
correctly (F displayed for front; R displayed for rear). Note that you must use either the front terminals
or rear terminals. You cannot mix connections.
Both front-panel and rear-panel connections are safety banana jacks. You can make these
connections with two insulated banana cables (for example, the leads included with the Keithley
Instruments Model 1756 General Purpose Test Lead Kit).
Connect the test leads as shown in the following figure.
10-16
Model DMM7510 7½ Digit Multimeter User's Manual