Section 10: Capturing and analyzing waveforms
Inductor current measurement device connections
You can use either the front or rear inputs. Ensure that the front-panel TERMINALS switch is set
correctly (F displayed for front; R displayed for rear). You must use either the front terminals or rear
terminals. You cannot mix connections.
Both front-panel and rear-panel connections are safety banana jacks. You can make these
connections with two insulated banana cables (for example, the leads included with the Keithley
Instruments Model 1756 General Purpose Test Lead Kit).
Connect the test leads as shown in the following figure.
Run the test using the front panel
Running the inductor current linearity test from the front panel:
1. Press the POWER button on the front panel to turn on the instrument.
2. On the FUNCTIONS swipe screen, select Digi I to select the digitize current function.
3. Swipe to the SETTINGS swipe screen.
4. Set the Sample Rate to 500,000.
5. Set the Count to 50.
6. Press the MENU key.
7. Select Graph to view the graph.
10-12
Figure 43: Inductor current linearity test connections
Model DMM7510 7½ Digit Multimeter User's Manual
DMM7510-900-01 Rev. C / September 2019