8.3

Storage

Keep the Finometer® Model-2 device in a clean environment to
protect it from dust and moisture.
See chapter 9 for
8
.4

Pressure calibration check

Pressure calibration should be checked at least once a year or when
there is doubt about the validity of the finger cuff pressure readings.
To check the pressure c
1. Connect the Finometer® Model-2 units as described in chapter 1.
2. Connect a calibrated precision manometer to air outlet C1 of the
Frontend Unit.
3. Press the
MARKER
calibration check mode while the syst
d
isplay shows the following message:
4.
C
heck zero level and then use the
p
ressure in steps of 50 to 250 mmHg. Use
t
he pressure in steps of 50 to 0 mmHg. At each pressure level the
p
recision manometer reading should be within 3 mmHg (1% full
s
cale) of the selected pressure level. When no precision manometer
i
s available, the check can be performed as follows:
• Close the air outlet C1 of the Frontend Unit with a finger
• Follow the pressure steps and selected pressure must be
displayed in value NNN on the Display.
5
.
Press the
START STOP
If the indicated pressures are not within tolerance, contact Finapres
Medical Systems B.V., [Customer support] to check the device.
User's Guide
environmental storage conditions.
alibration:
and
key simultaneously to enter the
PHYSIOCAL
Systolic Diastolic Mean
STEADY PRESSURE
G:NNN M:NNN mmHg
Physiocal
key to return to Ready mode.
/
em is in Ready mode. The LCD
Rate (bpm)
Height
key to increase the
PHYSIOCAL
key to decreas
HEIGHT
e

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Finometer midi

Table of Contents