44 Finometer® Model-2

4.2.2 Connect Frontend mode

When the Frontend Unit has not yet been connected, the display will
read, e.g.:
This display shows the message that the Frontend Unit is not
connected to the main unit. After connecting the Frontend Unit,
Finometer® Model-2 automatically switches to Ready mode.

4.2.3 Ready mode

After powering up Finometer® Model-2 with the Frontend Unit
connected, the control unit should display e.g.:
The following table describes the Ready mode display.
Item
Description
'BRA'
Beat-to-beat blood pressure data
derived from reconstructed
brachial artery pressure
'C1'
Finger cuff connection C1
-HHH
Height correction value in mmHg.
A negative sign means that the
Height Correction Unit transducer
(at the finger) is below heart
level; positive above.
User's Guide
Systolic Diastolic Mean
FINOMETER M2
CONNECT FRONTEND
Physiocal
Systolic Diastolic Mean
READY MODE:START
--
BRA C1 -HHH
Physiocal
Rate (bpm)
Height
Rate (bpm)
Height
Options
'FIN' Beat-to-beat blood
pressure data derived from
finger pressure
• value in range -128 to
+128 [mmHg].
• '
' if not connected.
NC
• '
' if not properly nulled
NZ
(zeroed).