5.1.1 Checking Height Correction Unit nulling

To check the cu
rrent Height Correction Unit nulling:
1. Keep the Height Correction Unit transducer an
the Height Correction Unit at the same level (Figure 5.1 Checking
the Height Correction Unit nulling).
Figure 5.1 Checking the Height Correction Unit nulling
2. Check the height value that is displayed in Ready mode on the LCD
display. This value should be '00'. Perform a Height Correction Unit
nulling procedure if this value is not '00'.
5.
1
.2 Performi
ng a Height Correction Unit nulling procedure
To perform a Height Correction Unit nulling procedure:
1
. The Finometer® Model-2 should be in Ready mode.
2. Keep the Height Correction Unit transducer and reference point
the Height Correction Unit at the same level (Figure 5.1 Checking
the Height Correction Unit nulling).
3. Press
the
HEIGHT
After about a second the following message should be displayed:
4
. The height value will be updated continuously.
5
. Press the
START
key to start the procedure.
Systolic Diastolic Mean
Heig
ht Null:STOP
OKAY
HEIGHT+HHH
Physiocal
key to return to Ready mode.
/
STOP