62 Finometer® Model-2
To obtain a good transmission of pressure from the airbladder to the
u
nd
erlying tissues both sides of the airbladder sh
(see Figure 6.1).
Cuff too large
Figure 6.1 Cuff size selection
6

.1.2 Warnings on cuff handling

Take into account the following notes to prevent finger cuff damage:
Don't remove the finger cuff from a finger before stopping the
measurement or before disconnecting the air hose from th
Unit. The finger cuff may be damaged even though Finometer® Model-
2 automatically takes pressure away when the finger cuff unwraps.
• Don't apply air pressure to a finger cuff, when it is not wrapped
around a finger or any other solid object! This may damage the
finger cuff.
• Don't bend finger cuffs outwards to a
damage the bonding and the electrical shielding. Finger cuffs are
preformed around a coni
Do not try to repair
this will su
bstantially affe

6.1.3 Cuff application

To wrap the finger cuff around the finger:
1. Bend the finger cuff open just enough to see the LE
2. Place the finger in the cuff s
symmetrically placed on each side of the finger's soft parts in the
center of the middle phalanx (Figure 6-2). The cuff cable and
User's Guide
Cuff ok!
Cuff too small
cal mandrel duri
defecti
ve finger cuffs wi
ct measu
rement accur
uch that the LED and photocell are
ould just make contact
e Frontend
flat shape since this may
ng manufacturi
ng.
th e.g. adhesi
acy.
D and photocell.