3.8

Connecting the Frontend Unit

To connect the Frontend Unit:
1. Apply the Frontend Unit to the subject's wrist and fasten the strap
so, that it cannot twist. It is recommended to use the non-
dominant arm.
2. Guide the Frontend Unit cable along the arm and fasten the arm
straps.
3. Connect the Frontend Unit connector to the Main Unit (connector
marked [Front end], see Figure 3.8, marked with D).
3.9

Connecting the Height Correction Unit

In addition to the above steps 1 through 3:
4. Fix the reference ending of the Height Correction Unit to the
subjects arm at heart level near the body. The Height Correction
Unit transducer should be fixed to the finger cuff. For more detailed
information about height level compensation see the Finger
Pressure Reference Guide.
3.10

Connecting the ECG (optional)

To connect the optional ECG module:
1. Connect the ECG module (section 3.5) to the main unit Lemo
connector marked ECG (connector marked [ECG], see Figure 3.8
marked with C).
2. Read the ECG Manual for further instructions on how to use the
ECG.
When the ECG module is connected to the Finometer® Model-2, the
combination complies with the IEC 60601-1-1 for electro medical
equipment.